Introduction
Step 1
Newport e500 Exhalation Module Disassembly
- Pull up and remove the flexible rubber exhalation outlet boot (push fit).
Pull up and remove the flexible rubber exhalation outlet boot (push fit).
1024
Step 2
- Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.
Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.
Step 3
- Pull the exhalation module straight up to remove it from the GDU.
Pull the exhalation module straight up to remove it from the GDU.
To reassemble your device, follow these instructions in reverse order.