Introduction
Note: When reinstalling flow sensors, note whether the sensor is labeled for air or oxygen and be sure to install in the appropriate position.
What you need
Step 1
Newport e500 Top Cover Disassembly
- Place device so that the top cover is facing you.
- Open the top cover.
Place device so that the top cover is facing you.
Open the top cover.
1024
Step 2
- Remove the four hex nuts and lock washers (two at each hinge) that secure the cover to the hinges on the left side of the pneumatic box.
Remove the four hex nuts and lock washers (two at each hinge) that secure the cover to the hinges on the left side of the pneumatic box.
Step 3
- Remove the cover and set aside. Retain hardware for re-assembly.
Remove the cover and set aside. Retain hardware for re-assembly.
Step 4
Newport e500 Exhalation Module Disassembly
- Pull up and remove the flexible rubber exhalation outlet boot (push fit).
Pull up and remove the flexible rubber exhalation outlet boot (push fit).
Step 5
- Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.
Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.
Step 6
- Pull the exhalation module straight up to remove it from the GDU.
Pull the exhalation module straight up to remove it from the GDU.
Step 7
Newport e500 Air and Oxygen Flow Sensors Disassembly
- Remove the 2 screws and washers securing the manifold/mixer assembly
Remove the 2 screws and washers securing the manifold/mixer assembly
Step 8
- Slide the manifold/mixer assembly forward to disengage the flow sensors
Slide the manifold/mixer assembly forward to disengage the flow sensors
Step 9
- Gently pull each flow sensor straight out from the servo valve, (the flow sensor push-fits into the servo valve port).
Gently pull each flow sensor straight out from the servo valve, (the flow sensor push-fits into the servo valve port).
Step 10
- Disconnect the flow sensor cable from its 10 pin connector at the underside of the flow sensor PCB and lift out the flow sensor.
Disconnect the flow sensor cable from its 10 pin connector at the underside of the flow sensor PCB and lift out the flow sensor.
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
Author
with 2 other contributors
Jeremy
Member since: 04/07/2020
589 Reputation
6 Guides authored
Badges:
5
+2 more badges
Team
Cal Poly, Team S7-G16, Paton Spring 2020
Member of Cal Poly, Team S7-G16, Paton Spring 2020
CPSU-PATON-S20S7G16
3 Members
13 Guides authored